FDA Adverse Event Malfunction Summary report: N

ATTAIN BIPOLAR OVER THE WIRE LEAD

MDR report key: 2790350 · Received October 15, 2012

Report

Report Number
2649622-2012-15016
Event Type
Malfunction
Date Received
October 15, 2012
Report Date
October 14, 2019
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015/S012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MET EXPECTED LONGEVITY.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS AT ELECTIVE REPLACEMENT INDICATOR (ERI) PREMATURELY. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE THRESHOLD WAS HIGH ON BOTH THE RIGHT VENTRICULAR (RV) AND LEFT VENTRICULAR (LV) LEADS, AND THERE WAS A POTENTIAL CONCERN ABOUT THE PERFORMANCE OF THE LV LEAD. DURING THE DEVICE CHANGE THE PHYSICIAN EVALUATED THE LEADS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED NOTED THE LEFT VENTRICULAR (LV) LEAD WAS TURNED OFF AND LATER CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN BIPOLAR OVER THE WIRE LEAD PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419488

Patients

Seq Age Sex Outcome Treatment
1 63 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD