CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2012-14124
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 30, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ES
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE LEAD WAS STRETCHED. IT WAS NOTED THAT ALL CONDUCTORS WERE STRETCHED, THE INNER INSULATION WAS KINKED/BUCKLED, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGED AT IMPLANT. VISUAL ANALYSIS NOTED THAT THE LEAD HAS BEEN STRETCHED CAUSING THE INNER TUBING TO BUCKLE. THIS PREVENTS PROPER TORQUE TRANSFER WHEN TURNING THE IS-1 PIN.
IT WAS REPORTED THAT DURING IMPLANT THERE WERE PROBLEMS WITH THE HELIX. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |