FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2790312 · Received October 15, 2012

Report

Report Number
2649622-2012-14124
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 30, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE LEAD WAS STRETCHED. IT WAS NOTED THAT ALL CONDUCTORS WERE STRETCHED, THE INNER INSULATION WAS KINKED/BUCKLED, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGED AT IMPLANT. VISUAL ANALYSIS NOTED THAT THE LEAD HAS BEEN STRETCHED CAUSING THE INNER TUBING TO BUCKLE. THIS PREVENTS PROPER TORQUE TRANSFER WHEN TURNING THE IS-1 PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT THERE WERE PROBLEMS WITH THE HELIX. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 Other