FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2790262 · Received October 15, 2012

Report

Report Number
2182208-2012-03438
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
May 21, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE REPORTED EVENT. THE BLANK SCREEN IS RELATED TO A PRINTED CIRCUIT BOARD ASSEMBLY THAT WAS FOUND OUT OF ELECTRICAL SPECIFICATION. ERRORS LOGGED ON THE HARD DRIVE ARE CONSISTENT WITH A PRINTED CIRCUIT BOARD ASSEMBLY CALIBRATION ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THE SCREEN WILL NOT COME ON. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER FOR REPAIR, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD