FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2790164 · Received October 15, 2012

Report

Report Number
2182208-2012-03428
Event Type
Malfunction
Date Received
October 15, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THERE WAS INTERMITTENT TELEMETRY DUE TO A LOOSE RADIOFREQUENCY (RF) HEAD CONNECTOR ON THE PROGRAMMER PRINTED CIRCUIT BOARD. IT WAS ALSO NOTED THAT THE KEYBOARD CABLE WAS DAMAGED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER "TAKES TOO LONG" TO DOWNLOAD FROM THE DEVICE. THE PROGRAMMER HEAD WAS CHANGED OUT BUT THE SITUATION DID NOT IMPROVE. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD| 2290 PACING SYSTEM ANALYZER