FDA Adverse Event
Injury
Summary report: N
CONCERTO CRT-D DR
MDR report key: 2790163
·
Received October 15, 2012
Report
- Report Number
- 6000144-2012-05309
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD |