FDA Adverse Event Injury Summary report: N

MAXIMO VR

MDR report key: 2790149 · Received October 15, 2012

Report

Report Number
6000094-2012-02251
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S37
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY : (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION, HOWEVER PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS REVEALED A POWER ON RESET OF THE PARAMETERS DUE TO A WRITE TO LOCKED RAM, ADDR=10C8, DATA=2 ON (B)(6) 2012 03:45:39. ADDITIONALLY THERE WAS ONE PATIENT ALERT FOR POR ON (B)(6) 2012 02:45:39.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ELECTRICAL RESET OCCURRED WITH THE DEVICE. THE DEVICE WAS INTERROGATED TO OBTAIN PERFORMANCE DATA, WHICH CLEARED THE ELETRICAL RESET. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7232CX

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention