FDA Adverse Event Malfunction Summary report: N

EXTERNAL PULSE GENERATOR

MDR report key: 2790140 · Received October 15, 2012

Report

Report Number
2183613-2012-01652
Event Type
Malfunction
Date Received
October 15, 2012
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THE RING WAS BENT, THE BATTERY DRAWER WAS BROKEN, AND THE KEYBOARD PAD WAS COSMETICALLY DAMAGED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR TESTING AND CALIBRATION. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Other