CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2012-14104
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 27, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE U.S. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THE PROXIMAL CONDUCTOR WAS STRETCHED, THE LEAD APPEARED DAMAGED AT IMPLANT AND THE LEAD WAS STRETCHED.
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEAD HELIX TIP "JUMPED" OUT OF THE LEAD WHEN USING THE ROTATION TOOL AND THEN THE HELIX TIP WOULD NOT RETRACT BACK INTO LEAD. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |