FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2790088 · Received October 15, 2012

Report

Report Number
2649622-2012-14103
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 25, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX WAS DISTORTED/BENT. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVEHEAD, THE TURNS TO EXTEND/RETRACT THE HELIX EXCEEDS SPECIFICATION, AND THE LEAD WAS DAMAGED AT IMPLANT. THE ANALYST NOTED THAT THE LEAD WAS RECEIVED WITH THE HELIX EXTENDED, STRETCHED, AND BENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT ONE DAY POST IMPLANT, THE LEAD HAD NO CAPTURE AND WAS DISLODGED. DURING THE LEAD REVISION, IT WAS NOT POSSIBLE TO SCREW THE LEAD INTO THE MYOCARDIUM. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| R