FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2790010 · Received October 15, 2012

Report

Report Number
2649622-2012-14692
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS STRETCHED, THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A POSSIBLE MYOCARDIAL PERFORATION THAT OCCURRED, FURTHER INFORMATION WAS UNABLE TO BE OBTAINED REGARDING THE PERFORATION, CAUSE, AND/OR INTERVENTION. HIGH THRESHOLDS WERE OBSERVED IN THE LEFT VENTRICULAR LEAD AND THE LEAD WAS REPORTED TO BE "COMPROMISED" DURING A "PAC CHANGE". IT WAS FURTHER REPORTED THAT OVERSENSING AND NOISE WERE ALSO OBSERVED IN THE RIGHT VENTRICULAR LEAD. THE LV LEAD WAS EXPLANTED AND REPLACED. THE RV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB