FDA Adverse Event Malfunction Summary report: N

EXTERNAL PULSE GENERATOR

MDR report key: 2790008 · Received October 15, 2012

Report

Report Number
2182208-2012-03416
Event Type
Malfunction
Date Received
October 15, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. THE EVENT OCCURRED OUTSIDE THE US. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) ANALYSIS FOUND THE LCD (LIQUID CRYSTAL DISPLAY) WAS OUT OF SPECIFICATION, THE VENTRICULAR OUTPUT CONNECTOR WAS CONTAMINATED, AND ONE SIDE BAIL COVER WAS BROKEN.

Description of Event or Problem · 1

THE EPG (EXTERNAL PULSE GENERATOR) WAS RETURNED TO THE MANUFACTURER FOR A DEVICE CHECK. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTERNAL PULSE GENERATOR DTE MEDTRONIC, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Other