FDA Adverse Event Injury Summary report: N

LOOK IRRIGATING ANTERIOR CHAMBER CANNULA 30G

MDR report key: 278605 · Received May 17, 2000

Report

Report Number
1643219-2000-00001
Event Type
Injury
Date Received
May 17, 2000
Date of Event
August 13, 1999
Report Date
May 17, 2000
Manufacturer
SURGICAL SPECIALTIES CORP
Product Code
HMX
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SURGEON'S INITIAL REPORT STATED THE CANNULA BLEW OFF THE TIP OF SYRINGE AND CREATED A LARGE IRIDODIALYSIS, ZONULAR RUPTURE WITH IMMEDIATE COPIOUS HEMORRHAGING, VITRECTOMY PERFORMED TO ATTEMPT TO REPAIR IRIS. IRIS WAS TOO BROKEN UP TO SUTURE. INTRAOCULAR LENS RETRIEVED WITH FORCEPS. IN SURGEON'S LETTER TO RETINAL SPECIALIST, SURGEON STATED CANNULA SEPARATED FROM SYRINGE AND EITHER THE CANNULA OR A STRONG STREAM OF BSS (BALANED STERILE SALINE) ENTERED THE EYE, FORMED AN IRIDODIALYSIS TEMPORALLY, LACERATED THE IRIS, DISPLACED THE IOL (INTRAOCULAR LENS) AND TORE ZONULES AND THE CAPSULE. VERBAL COMMUNICATION FROM THE INITIAL REPORTER STATED THAT ADDITIONAL SURGERY WAS PERFORMED ON 10/12/1999. THE POST-OPERATIVE EVAL ON 11/8/1999 REVEALED THAT THE PT HAD NO SIGHT IN THE RIGHT EYE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LOOK IRRIGATING ANTERIOR CHAMBER CANNULA 30G IRRIGATING ANTERIOR CHAMBER CANNULA 30G HMX SURGICAL SPECIALTIES CORP 3405 MJ61420

Patients

Seq Age Sex Outcome Treatment
1 76 YR Disability