FDA Adverse Event Malfunction Summary report: N

HFV PATIENT CIRCUIT

MDR report key: 2784981 · Received September 14, 2012

Report

Report Number
1719232-2012-00001
Event Type
Malfunction
Date Received
September 14, 2012
Date of Event
September 10, 2012
Report Date
September 14, 2012
Manufacturer
BUNNELL
Product Code
LSZ
PMA / PMN Number
P850064
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE RRT CALLED TO REPORT THAT TWO CIRCUITS HAD OVERHEATED AND ONE OF THEM HAD MELTED THE HEATER WIRE INSULATION AND CIRCUIT TUBE. BOTH FAILURES OCCURRED WITH THE SAME PT. THE PT WAS NOT COMPROMISED AND CONTINUED TO BE VENTILATED WITH THE SAME HIGH FREQUENCY VENTILATOR (HFV) AND A DIFFERENT PT CIRCUIT. A BUNNELL REP VISITED THE FACILITY ON (B)(4) 2012 AND VISUALLY CONFIRMED THE FAILURE ON ONE CIRCUIT. VISUALLY, THE OTHER CIRCUIT APPEARED INTACT AND UNDAMAGED. THE HIGH FREQUENCY VENTILATOR WAS ALSO CHECKED AND WAS OPERATING NORMALLY. BUNNELL INITIATED AN INTERNAL CORRECTIVE ACTION REQUEST (CAR), NUMBER (B)(4), ON (B)(4) 2012. A SUPPLEMENTAL PMA, P850064/S021, WAS SUBMITTED ON (B)(4) 2012 TO INCREASE THE HEATER WIRE INSULATION TEMP RATING WHICH WILL SIGNIFICANTLY REDUCE THE PROBABILITY OF THIS TYPE OF RARE FAILURE OCCURRING. ADDITIONAL INFO RECEIVED FROM THE FACILITY, FAILURE INVESTIGATION OR FROM THE CAR WILL BE ADDED TO THIS REPORT.

Description of Event or Problem · 1

RESPIRATORY CARE DEPT CALLED TO REPORT THAT TWO CIRCUITS HAD OVERHEATED AND ONE OF THEM HAD MELTED THE HEATER WIRE INSULATION AND CIRCUIT TUBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HFV PATIENT CIRCUIT CIRCUIT LSZ BUNNELL 902 12E224

Patients

Seq Age Sex Outcome Treatment
1