FDA Adverse Event
Malfunction
Summary report: N
TAP III ELITE
MDR report key: 2783993
·
Received October 1, 2012
Report
- Report Number
- 3003496134-2012-00021
- Event Type
- Malfunction
- Date Received
- October 1, 2012
- Date of Event
- September 26, 2012
- Report Date
- September 26, 2012
- Manufacturer
- AIRWAY MANAGEMENT, INC.
- Product Code
- LRK
- PMA / PMN Number
- K062951
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- DENTIST
Narratives
Description of Event or Problem · 1
PATIENT REPORTED TO HIS DENTIST THAT THE HOOK OF HIS ORAL APPLIANCE HAD BROKEN LOOSE FROM THE UPPER TRAY. THERE WAS NO HARM TO THE PATIENT. INVESTIGATION SHOWED EVIDENCE THAT THE PATIENT HAD BITTEN DOWN ONTO THE LOWER TRAY, PUTTING EXCESSIVE FORCE ON THE HOOK. THE REPLACEMENT DEVICE FABRICATED AND SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAP III ELITE | SLEEP APNEA DEVICE | LRK | AIRWAY MANAGEMENT, INC. | TAP3100-0BMQ-12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |