FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2782175 · Received October 11, 2012

Report

Report Number
2649622-2012-12914
Event Type
Injury
Date Received
October 11, 2012
Date of Event
May 22, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS FOUND TO BE DISLODGED DURING A FOLLOW UP VISIT. THE LEAD WAS CAPPED AND REPLACED. DURING THE LEAD REVISION IT WAS IMPOSSIBLE TO FASTEN THE LEAD IN THE PACEMAKER DUE TO THE SETSCREW NOT WORKING AS IT SHOULD. THE PACEMAKER WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 20 YR Hospitalization| R 4076 IMPLANTABLE PACING LEAD