FDA Adverse Event Injury Summary report: N

HOVERMATT LATERAL PATIENT TRANSFER DEVICE

MDR report key: 2781990 · Received September 18, 2012

Report

Report Number
2531468-2012-00002
Event Type
Injury
Date Received
September 18, 2012
Date of Event
April 26, 2012
Report Date
September 17, 2012
Manufacturer
D.T. DAVIS ENTERPRISES LTD. T/A HOVERTECH INTERNATIONAL
Product Code
FRZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AS STATED IN THE REPORT FROM (B)(6): "AFTER THE INITIAL INVESTIGATION IT APPEARS THAT EITHER PROPER PT TRANSFER PROTOCOLS WHICH REQUIRE HOSPITAL BED BRAKES TO BE APPLIED BEFORE A TRANSFER COMMENCES WERE NOT ADHERED TO, OR THAT THE BRAKES ON THE BED WERE FAULTY." IT ALSO STATES IN THE HOVERMATT INSTRUCTIONS FOR USE; "ENSURE THE TRANSFER SURFACES ARE AS CLOSE AS POSSIBLE, AND BRAKE THE WHEELS." CONCLUSION: IT WOULD APPEAR THAT EITHER PROPER PT TRANSFER PROTOCOLS WHICH REQUIRE THE HOSPITAL BED BRAKES TO BE APPLIED BEFORE A TRANSFER COMMENCES WERE NOT ADHERED TO, OR THAT THE BRAKES ON THE BED WERE FAULTY. MY INVESTIGATION ALSO FOUND THAT THE HOVERMATT WAS OPERATING CORRECTLY AT THE TIME OF THE INCIDENT AND WAS INFLATED FIRMLY THOUGH TO SUPPORT THE PT AS THEY FELL. IT WAS ALSO SAID THAT THE HOVERMATT HAD IN FACT SAVED THE PATENT FROM RECEIVING A MUCH WORSE INJURY. IF THE PT HAD BEEN ON ANY OTHER TRANSFER DEVICE, THEY WOULD HAVE FALLEN DIRECTLY ONTO THE FLOOR.

Description of Event or Problem · 1

(B)(6) WAS DROPPED TO THE FLOOR WHILE BEING TRANSFERRED FROM HOSPITAL BED TO X-RAY TABLE 3 DAYS AFTER HAVING BILATERAL HIP REPLACEMENT. COULD HAVE BEEN OPERATOR ERROR OR EQUIPMENT FAILURE. DAMAGE TO LEFT HIP RESULTING IN BEING IN HOSPITAL A MONTH LONGER THAN EXPECTED, AND NEEDING TWO EXTRA OPERATIONS." THE EQUIPMENT FAILURE MENTIONED ABOVE IS THAT OF THE HOSPITAL BED BRAKES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOVERMATT LATERAL PATIENT TRANSFER DEVICE DEVICE, PATIENT TRANSFER POWERED FRZ D.T. DAVIS ENTERPRISES LTD. T/A HOVERTECH INTERNATIONAL HM34HS NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization