FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2781971
·
Received October 5, 2012
Report
- Report Number
- 8020893-2012-00864
- Event Type
- Malfunction
- Date Received
- October 5, 2012
- Date of Event
- September 7, 2012
- Report Date
- September 7, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCO
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) AND BACKLIGHT INVERTER PCBS. THE UNIT PASSED EXTENDED SELF TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCO | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |