FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2781971 · Received October 5, 2012

Report

Report Number
8020893-2012-00864
Event Type
Malfunction
Date Received
October 5, 2012
Date of Event
September 7, 2012
Report Date
September 7, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) AND BACKLIGHT INVERTER PCBS. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCO 840

Patients

Seq Age Sex Outcome Treatment
1