FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 2781729 · Received October 11, 2012

Report

Report Number
2649622-2012-12894
Event Type
Injury
Date Received
October 11, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE DEFIBRILLATION COIL WAS FRACTURED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WITH BILUMEN TUBING VOIDS, THE INNER INSULATION WAS BREACHED, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE OUTER INSULATION WAS BREACHED AND HAD A DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED AFTER BEING EXPLANTED DUE TO AN ELECTIVE REPLACEMENT. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6943

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R 6937 IMPLANTABLE TACHY LEAD