FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 2781727 · Received October 11, 2012

Report

Report Number
2649622-2012-12893
Event Type
Injury
Date Received
October 11, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, WAS ANALYZED AND ANALYSIS REVEALED A BREACHED DEPRESSION OF THE OUTER INSULATION. IT WAS NOTED THE DEFIBRILLATION CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION AND A COSMETIC DEPRESSION.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO AN UNRELATED INFECTION, WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6945

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB