MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2012-00532
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- May 24, 2012
- Report Date
- January 10, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K012833
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4).
THE INSTRUMENT WAS RETURNED AND EVALUATED. CUSTOMER INITIALLY REPORTED FAULTY INSTRUMENT. HOWEVER FOR CLARIFICATION, THE DEFECT WAS A BROKEN PITCH CABLE. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. IDLER PULLEY WAS FOUND TO BE PHYSICALLY DAMAGED. IDLER PULLEY EXHIBITED INDENTATIONS AND RIDGES INDICATIVE OF MISHANDLING AND MAY HAVE CAUSED PITCH CABLE TO BREAK. TUBE INSULATION ALSO APPEARS TO HAVE SCRATCHES AND GOUGE MARKS WITH THE SURFACE REMOVED. EVIDENCE NOT CONCLUSIVE, BUT DAMAGE IS LIKELY CAUSED BY MISHANDLING/MISUSE. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: - PROPER CARE AND HANDLING IS ESSENTIAL FOR SATISFACTORY PERFORMANCE OF SURGICAL INSTRUMENTS AND ACCESSORIES. EXAMINE THE INSTRUMENT OR ACCESSORY - INCLUDING ALL OF ITS COMPONENTS - THOROUGHLY BEFORE AND AFTER EACH USE. IF ANY ABNORMALITY IS FOUND, DO NOT USE IT. USE THE DEVICE FOR ITS INTENDED PURPOSE ONLY. - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. - DO NOT USED A DAMAGED CANNULA OR REDUCER. DAMAGED CANNULAE AND REDUCERS MAY ABRADE THE INSTRUMENT SHAFT AND GENERATE PARTICLES THAT MAY FALL INSIDE THE PATIENT.
THE CUSTOMER REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS FOUND FAULTY. THE INSTRUMENT WAS REPLACED WITH ANOTHER MARYLAND BIPOLAR FORCEPS INSTRUMENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2012 ISI RECEIVED ADDITIONAL INFORMATION INDICATING A PATIENT X-RAY WAS PERFORMED POST OPERATIVELY. THE X-RAY SHOWED NO FOREIGN BODY AND THE PATIENT DID NOT DEVELOP ANY COMPLICATIONS AFTER THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10110706 942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Other | DA VINCI SI SURG SYSTEM, INSTRUMENTS & ACCESSORIE |