FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2781645 · Received October 10, 2012

Report

Report Number
6000144-2012-04721
Event Type
Injury
Date Received
October 10, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE HAD NORMAL BATTERY DEPLETION AND MET 95% OF EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD POSSIBLE EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154AWG

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD