FDA Adverse Event Malfunction Summary report: N

5 MM MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 2781630 · Received October 10, 2012

Report

Report Number
2955842-2012-00518
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE MAIN TUBE INSULATION EXHIBITED DAMAGE ALONG THE ENTIRE MAIN TUBE. THE INSULATION WAS FOUND WITH SEVERAL GOUGE MARKS. THE MARKS WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. MAIN TUBE ALSO EXHIBITED A COUPLE OF DIFFERENT DAMAGED SECTIONS WITH TUBE INSULATION REMOVED. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. INTUITIVE SURGICAL CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT. IT WAS INDICATED THAT NO PIECES FELL INSIDE THE PATIENT DURING THE LAST SURGICAL PROCEDURE THIS INSTRUMENT WAS USED. THE PATIENT HAS NOT RETURNED TO THE HOSPITAL WITH ANY POST-SURGICAL COMPLICATIONS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI TRANSANAL TUMOR RESECTION SURGERY, THE CUSTOMER REPORTED THAT 'THE BLACK COATING MAIN INSULATION WAS COMING OFF' ON THE MONOPOLAR CAUTERY INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 5 MM MONOPOLAR CAUTERY INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420142-04 S10110415 126

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SURG SYSTEM, INSTRUMENTS & ACCESSORIES