FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2781614 · Received October 10, 2012

Report

Report Number
2955842-2012-00519
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN NEAR THE PROXIMAL PULLEYS AND PROXIMAL CLEVIS CABLE HOLE. THE IDLER PULLEY SPINS FREELY AND DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STICKS OUT AT WRIST. PROXIMAL CLEVIS CABLE HOLE EXHIBITED WEAR ON ONE EDGE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A VAGINAL CUFF CLOSURE IN THE COURSE OF A DAVINCI S HYSTERECTOMY PROCEDURE, THE SURGEON NOTICED THAT THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT DID NOT HAVE ANY GRIP STRENGTH; UPON FURTHER INSPECTION THE TEAM NOTICED THAT ONE OF THE INTERNAL CABLES WAS BROKEN. THE INSTRUMENT WAS EXCHANGED AND THE SURGICAL PROCEDURE WAS COMPLETED AS INTENDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10110803 057

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI S SURG SYSTEM, INSTRUMENTS & ACCESSORIES