FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2781601
·
Received October 10, 2012
Report
- Report Number
- 3004209178-2012-90010
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- September 28, 2012
- Report Date
- September 28, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP RECEIVED WITH PRIME ALARM DUE TO CRACKED END CAP.
Description of Event or Problem · 1
THE CUSTOMER REPORTED HAVING HIGH BLOOD GLUCOSE OF 250MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER MENTIONED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE MANUAL PRIME. THE CUSTOMER ALSO MENTIONED THAT THE MOTOR PUSHED OUT THE BACK PANEL OF THE INSULIN PUMP AND THERE IS DAMAGE ALSO. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |