FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 2781597
·
Received October 10, 2012
Report
- Report Number
- 3004209178-2012-89999
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- February 3, 2011
- Report Date
- October 5, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- K031390
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP POWER UP PROPERLY AFTER THE BATTERY WAS INSTALLED. UNABLE TO CONFIRM MISSING SEGMENTS. HOWEVER, THE DEVICE FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD PARTIAL DISPLAY. IT WAS STATED THAT THE BATTERY WAS CHANGED, BUT THE ANOMALY PERSISTED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM INSULIN INFUSION PUMP | LZG | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-712WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |