FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2781597 · Received October 10, 2012

Report

Report Number
3004209178-2012-89999
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
February 3, 2011
Report Date
October 5, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
K031390
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP POWER UP PROPERLY AFTER THE BATTERY WAS INSTALLED. UNABLE TO CONFIRM MISSING SEGMENTS. HOWEVER, THE DEVICE FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD PARTIAL DISPLAY. IT WAS STATED THAT THE BATTERY WAS CHANGED, BUT THE ANOMALY PERSISTED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP LZG OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712WWL

Patients

Seq Age Sex Outcome Treatment
1