FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2781582 · Received October 10, 2012

Report

Report Number
2024168-2012-06421
Event Type
Injury
Date Received
October 10, 2012
Date of Event
August 28, 2012
Report Date
September 17, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT OF A CUFF MISS WAS NOT CONFIRMED. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE ANTERIOR CUFF WAS CAPTURED AT SOME POINT DURING NEEDLE DEPLOYMENT PULLING THE LINK OUT OF THE POSTERIOR CUFF. THE LINK TO CUFF DETACHMENT WOULD APPEAR SIMILAR TO THE REPORTED CUFF MISS. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS RETRACTED A CUFF MISS HAD OCCURRED. THE VESSEL WAS RE-WIRED AND THE PROGLIDE DEVICE WAS REMOVED. HEMOSTASIS WAS ACHIEVED USING A SECOND PROGLIDE DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 20721J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention