PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-06413
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 17, 2012
- Report Date
- September 18, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT OF A CUFF MISS WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED NO SUTURE WAS PRESENT. THE METHOD USED TO ACHIEVE HEMOSTASIS WAS NOT REPORTED. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, THE FOLLOWING ADDITIONAL INFORMATION INDICATES THAT THE CUFF MISS OCCURRED DURING ATTEMPTED SUTURE PLACEMENT USING THE PRECLOSE TECHNIQUE. A SECOND PROGLIDE DEVICE WAS USED FOR SUCCESSFUL SUTURE PLACEMENT USING THE PRECLOSE TECHNIQUE. THE ARTERIOTOMY WAS A 6F. THE SHEATH WAS UPSIZED TO A 12F AND THEN 18F. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20627J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention | SHEATH: 6FR, 12FR, 18FR| HEPARIN |