FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2781517 · Received October 10, 2012

Report

Report Number
2024168-2012-06405
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 12, 2012
Report Date
September 18, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE WAS DISCARDED AND IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. REVIEWS OF THE LOT HISTORY RECORD AND COMPLAINT HISTORY OF THE REPORTED LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. REPORTEDLY THE COMMON FEMORAL ARTERY WAS MODERATELY CALCIFIED. THE INSTRUCTIONS FOR USE STATE UNDER SPECIAL PATIENT POPULATIONS THAT THE SAFETY AND EFFECTIVENESS OF THE PERCLOSE PROGLIDE SMC DEVICES HAVE NOT BEEN ESTABLISHED IN PATIENT POPULATIONS WITH FEMORAL ARTERY CALCIUM, WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER PERCLOSE PROGLIDE DEVICES, ARE BEING FILED UNDER A SEPARATE MEDWATCH MANUFACTURER REPORT NUMBERS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PROCEDURE, PRE-CLOSE PLACEMENT OF THE SUTURES WAS ATTEMPTED IN A MODERATELY CALCIFIED COMMON FEMORAL ARTERY USING PERCLOSE PROGLIDE DEVICES. REPORTEDLY, THE DEVICES WERE INSERTED INTO THE VESSEL THROUGH A 6-FRENCH SIZED ACCESS SITE. DURING EACH ATTEMPT TO DEPLOY THE SUTURES OF THREE PROGLIDES DEVICES, A NEEDLE-TO-CUFF MISS OCCURRED. WHEN THE NEEDLE PLUNGER WAS REMOVED OF EACH DEVICE, NO SUTURE WAS PRESENT ON THE NEEDLES. TWO ADDITIONAL PROGLIDE DEVICES WERE SEQUENTIALLY DEPLOYED 60-DEGREES OPPOSITE IN ORIENTATION AND SET TO THE SIDE. THE ACCESS SITE WAS DILATED TO 18-FRENCH TO MATCH THE OUTER DIAMETER OF THE DELIVERY CATHETER. AFTER CONCLUSION OF THE TAVI PROCEDURE, THE KNOTS FROM THE TWO PROGLIDE DEVICES WERE SEQUENTIALLY ADVANCED AND TIGHTENED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE PHYSICIAN WAS REPORTED TO BE IS REPORTEDLY TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATHS: 6-FRENCH