FDA Adverse Event Malfunction Summary report: N

MEGA NEEDLE DRIVER INSTRUMENT

MDR report key: 2781503 · Received October 10, 2012

Report

Report Number
2955842-2012-00511
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 4, 2012
Report Date
September 4, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K082497
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT WRIST. THE OTHER CABLES AT WRIST ARE NOT DAMAGED. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE WIRE ON THE MEGA NEEDLE DRIVER INSTRUMENT FRAYED AT THE DISTAL END AND SPLIT IN HALF. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. LATE SUBMISSION OF THIS REPORT IS DUE TO AN OVERSIGHT BY THE RESPONSIBLE COMPLAINT HANDLER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGA NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420194-08 M10120724 820

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM & ACCESSORIES