PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-06398
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 20, 2012
- Report Date
- September 20, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS NOT CONFIRMED BECAUSE KEY COMPONENTS WERE NOT RETURNED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE MODERATELY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PERCLOSE PROGLIDE DEVICE AFTER A CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, DURING KNOT ADVANCEMENT THE ENTIRE SUTURE CAME OUT OF THE PATIENT'S ANATOMY. A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS A 20 MINUTE DELAY IN PROCEDURE. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20118J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention | SHEATH: 6F ANGIOMAX |