FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 2781303 · Received October 10, 2012

Report

Report Number
9680353-2012-00077
Date Received
October 10, 2012
Date of Event
September 10, 2012
Report Date
October 10, 2012
Manufacturer
STERIS CANADA CORPORATION
Product Code
LDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE STERIS SERVICE TECHNICIAN INSPECTED THE UNIT AND FOUND A BREAK IN THE DRYER PISTON ASSEMBLY, ALLOWING WATER TO LEAK FROM THE WASHER THROUGH THE VENT HOSE. THE TECHNICIAN REPLACED THE BROKEN PARTS, RAN TEST CYCLES TO CONFIRM THE UNIT WAS OPERATING PROPERLY AND RETURNED IT TO SERVICE.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A RELIANCE SYNERGY WASHER-DISINFECTOR LEAKED WATER ONTO THE FLOOR EXTENDING APPROXIMATELY 6 FEET FROM THE WASHER LOAD SIDE. NO INJURIES, PROCEDURE CANCELLATIONS OR WATER DAMAGE WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RELIANCE SYNERGY WASHER WASHER LDS STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1