FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2781259 · Received October 10, 2012

Report

Report Number
1416980-2012-02027
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
August 13, 2012
Report Date
September 19, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR; TIDAL TOTAL ULTRA-FILTRATION (UF) REMOVAL WAS SET TOO LOW. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012 22:57:41. DURING NIGHT DRAIN CYCLE 11, THE PATIENT'S ULTRAFILTRATION READING WAS 2198ML, INDICATING THE HOME PATIENT (HP) DRAINED 1398ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1