FDA Adverse Event Injury Summary report: N

FLEXTEND II

MDR report key: 2781029 · Received October 10, 2012

Report

Report Number
2124215-2012-12946
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 24, 2012
Report Date
November 14, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE LEAD HAS BEEN RETURNED AND ANALYSIS COMPLETED, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. ONLY THE DISTAL SEGMENT OF THE LEAD WAS RETURNED. THE ELECTRODE END HAD BLOOD FILTRATION IN THE HELIX HOUSING. THE LEAD TIP WAS ALSO FOUND TO BE BENT AT A 6 DEGREE ANGLE BETWEEN THE HELIX HOUSING AND ANODE RING. FURTHER VISUAL INSPECTION NOTED ABRASIONS ON THE LEAD'S INSULATION IN TWO LOCATIONS, INDICATIVE OF LEAD-ON-LEAD CONTACT . AN X-RAY WAS COMPLETED WHICH CONFIRMED THAT THE LEAD'S CONDUCTOR COIL WAS FRACTURED IN ONE OF THE AREAS WHERE THE INSULATION WAS DAMAGED . IT WAS ALSO NOTED THAT THE LEAD SUTURE SLEEVE WAS CLOSE TO THE FRACTURED COIL. LABORATORY ANALYSIS CONFIRMED THE CLINICAL OBSERVATIONS OF NOISE, OVERSENSING, INSULATION DAMAGE AND THE LEAD FRACTURE. BASED ON THE ANALYSIS RESULTS, WE SUSPECT THAT FATIGUE OR STRESS IN THE REGION OF THE FRACTURE SITE DUE TO RELATIVE MOTION BETWEEN THE SUTURE SLEEVE AND AN ANATOMICAL FEATURE LED TO THE FRACTURE. AS A LEAD MOVES IN RESPONSE TO NORMAL HEART RHYTHMS AND BLOOD FLOW, EXTENSIVE FLEXING OVER A PERIOD OF TIME MAY CAUSE FATIGUE OR STRESS, WEAKENING THE COIL, ULTIMATELY RESULTING IN A FRACTURE. THIS CAN OCCUR BETWEEN THE SUTURE SLEEVE AND SOME OTHER PART OF THE ANATOMY. FATIGUE FRACTURES IN THE POCKET OR CLAVICULAR/FIRST RIB AREA ARE WELL KNOWN AND DOCUMENTED IN THE INDUSTRY. A COMBINATION OF LEAD DESIGN, IMPLANT TECHNIQUES, AND PATIENT ANATOMY AND ACTIVITY LEVEL CONTRIBUTE TO THESE TYPES OF OCCURRENCES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE THAT WAS OVERSENSED AND RESULTED IN THE DELIVERY OF INAPPROPRIATE SHOCKS. A LEAD FRACTURE WAS CONFIRMED. DURING THE REVISION PROCEDURE, SEVERAL INSULATION BREACHES WERE NOTED ON THIS LEAD IN THE POCKET REGION. THE PHYSICIAN SUSPECTS THAT LEAD ON LEAD CONTACT CAUSED THE INSULATION ABRASIONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE RV LEAD WAS EXPLANTED AND WILL BE RETURNED FOR LABORATORY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4097

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| L| R T175| 4097| 1793| 1741| 0049| 0125