EASYTRAK 3
Report
- Report Number
- 2124215-2012-13185
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 27, 2012
- Report Date
- October 26, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES.LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT ONE DAY POST IMPLANT OF THIS LEAD, THE LEAD WAS NOTED TO BE DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE.
THE LEAD WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASYTRAK 3 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4549 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |