ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2012-13330
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 19, 2012
- Report Date
- October 26, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
THIS RIGHT VENTRICULAR (RV) LEAD IS EXPECTED TO BE RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RECEIPT OF ADDITIONAL INFORMATION, OR WHEN ANALYSIS IS COMPLETE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD DISLODGED AND SENSED ATRIAL SIGNALS, WHICH LED TO AN INAPPROPRIATE SHOCK. NOISE WAS ALSO NOTED, AS WELL AS LOSS OF CAPTURE. THE LEAD DISLODGEMENT WAS CONFIRMED BY AN X-RAY. THE PATIENT WAS NOT DEPENDENT ON PACING THERAPY, AND THERE WERE NO ADVERSE PATIENT EFFECTS. IT WAS NOTED THAT WHEN THIS LEAD HAD BEEN IMPLANTED IN (B)(6) 2012, IT WAS IMPLANTED IN THE APICAL POSITION AND THERE WERE EIGHT ATTEMPTS TO POSITION THIS LEAD TO FIND GOOD MEASUREMENTS. IT WAS THOUGHT THAT THE HELIX MAY HAVE BEEN DAMAGED DURING THE REPOSITIONING ATTEMPTS AT THE IMPLANT PROCEDURE. THE PATIENT WAS HOSPITALIZED FOR A LEAD REVISION PROCEDURE. DURING THE REVISION PROCEDURE, THE LEAD WAS EASILY REMOVED BY SIMPLY PULLING IT OUT. A NEW RV LEAD WAS IMPLANTED IN THE SEPTAL POSITION WITH NO DIFFICULTIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0295 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |