EASYTRAK 2
Report
- Report Number
- 2124215-2012-13220
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 26, 2012
- Report Date
- October 16, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY THE LEAD WAS THOROUGHLY INSPECTED AND ANALYZED. ANALYSIS REVEALED ALL CONDUCTOR COILS ARE FRACTURED APPROXIMATELY 43.9 CENTIMETERS FROM THE TERMINAL PIN. THE FRACTURE APPEARED TO BE AT PROXIMAL END OF THE SUTURE SLEEVE TIE-DOWN AREA. THE CLINICAL OBSERVATIONS WERE ABLE TO BE CONFIRMED.
AS OF TODAY, THE LEAD HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
THE LEAD HAS BEEN RETURNED FOR ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR LEAD DISPLAYED HIGH, OUT OF RANGE PACE IMPEDANCE MEASUREMENTS. A LEAD FRACTURE WAS SUSPECTED. THE PATIENT WAS SEEN FOR A LEAD REVISION WHERE THE LEAD WAS EXPLANTED AND REPLACED. THE LEAD IS TO BE RETURNED FOR ANALYSIS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Hospitalization| L| R | 4470| 0185| 1290| H219| 4471| 4518 |