FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2781006 · Received October 10, 2012

Report

Report Number
2124215-2012-13220
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 26, 2012
Report Date
October 16, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY THE LEAD WAS THOROUGHLY INSPECTED AND ANALYZED. ANALYSIS REVEALED ALL CONDUCTOR COILS ARE FRACTURED APPROXIMATELY 43.9 CENTIMETERS FROM THE TERMINAL PIN. THE FRACTURE APPEARED TO BE AT PROXIMAL END OF THE SUTURE SLEEVE TIE-DOWN AREA. THE CLINICAL OBSERVATIONS WERE ABLE TO BE CONFIRMED.

Additional Manufacturer Narrative · 1

AS OF TODAY, THE LEAD HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

THE LEAD HAS BEEN RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR LEAD DISPLAYED HIGH, OUT OF RANGE PACE IMPEDANCE MEASUREMENTS. A LEAD FRACTURE WAS SUSPECTED. THE PATIENT WAS SEEN FOR A LEAD REVISION WHERE THE LEAD WAS EXPLANTED AND REPLACED. THE LEAD IS TO BE RETURNED FOR ANALYSIS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4518

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| L| R 4470| 0185| 1290| H219| 4471| 4518