CONTAK RENEWAL
Report
- Report Number
- 2124215-2012-13321
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- September 25, 2012
- Report Date
- November 8, 2012
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION WAS COMPLETED TO ASSESS THE RATE OF BATTERY DEPLETION. GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE, THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, DEFIBRILLATION, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
THIS DEVICE WILL BE EXPLANTED AND REPLACED IN THE NEAR FUTURE. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.
SUBSEQUENTLY, ADDITIONAL INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. THIS DEVICE WAS EXPLANTED AND REPLACED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI), AND THRESHOLD MEASUREMENTS HAVE INCREASED OVER TIME. THERE WAS CONCERN THE BATTERY DEPLETED EARLIER THAN EXPECTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.