FDA Adverse Event Malfunction Summary report: N

FLEXTEND

MDR report key: 2780963 · Received October 10, 2012

Report

Report Number
2124215-2012-13280
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 25, 2012
Report Date
October 26, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES

Additional Manufacturer Narrative · 1

AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE DEVICE CHANGE OUT PROCEDURE FOR NORMAL BATTERY DEPLETION, AN X-RAY REVEALED THAT THE RIGHT ATRIAL (RA) LEAD WAS DISLODGED. SUBSEQUENTLY THE EXISTING RA LEAD WAS EXPLANTED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. IT WAS NOTED THAT THE DEVICE HAD BEEN PROGRAMMED VVI AT AN UNDEFINED TIME IN THE PAST FOR AN UNKNOWN REASON. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4086

Patients

Seq Age Sex Outcome Treatment
1 52 YR 4086| 1861| 0185