OT VERIO IQ METER
Report
- Report Number
- 3008382007-2012-04940
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Report Date
- September 14, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (11/14/2012)-DEVICE EVALUATION: THE PRODUCTS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (PA) RESPECTIVELY ON (B)(4) 2012 AND (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE PRIMARY COMPLAINT WAS CONFIRMED, THE METER DID NOT RECOGNIZE THE CONTROL SOLUTION WHEN IT WAS APPLIED TO THE TEST STRIP. THE METER'S SPC PIN 2 WAS FOUND TO BE LOW. THE TEST STRIPS PASSED TESTING WITH NO FAULTS IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2012, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) IN (B)(4) ALLEGING A ONETOUCH VERIO IQ METER WAS DISPLAYING AN APPLY SAMPLE MESSAGE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION(S): THERE IS NO INDICATION THAT SUBJECT METER CAUSE OR CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY SYMPTOMS, TREATMENT, OR BLOOD GLUCOSE VALUES SUGGESTIVE OF A SERIOUS INJURY. HOWEVER THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3339371 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |