FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 2780807 · Received October 10, 2012

Report

Report Number
2939301-2012-11786
Event Type
Malfunction
Date Received
October 10, 2012
Report Date
October 1, 2012
Manufacturer
LIFESCAN INC.
Product Code
NBW
PMA / PMN Number
K073231
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT IS EXPECTED TO BE RETURNED.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE A PATIENT CONTACTED LIFESCAN ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH WHEN COMPARED TO A LAB READING. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "158 MG/DL" WITH A LIFESCAN METER AND "119 MG/DL" ON A LABORATORY DEVICE, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BETWEEN THE TESTS THERE WAS NO INTERVENTION THAT WOULD BE EXPECTED TO CHANGE THE BLOOD GLUCOSE. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR ACCURACY, GREATER THAN 1 MG/DL (0.056 MMOL/L) PER MINUTE AND GREATER THAN 20 MG/DL (1.11 MMOL/L) OR 20%. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC.

Patients

Seq Age Sex Outcome Treatment
1