FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2780731 · Received October 10, 2012

Report

Report Number
2124215-2012-12907
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 17, 2012
Report Date
October 12, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. A CUT NOTED IN THE INSULATION AT 35 CENTIMETERS FROM THE PIN WHICH LIKELY OCCURRED AT EXPLANT. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN ACUTE RISE IN PACING THRESHOLDS AND LOSS OF CAPTURE (LOC). THE LEAD WAS EXPLANTED AND A NEW RV LEAD SUCCESSFULLY IMPLANTED. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE NVY CPI - DEL CARIBE 0293

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| L| R E140| 0293