FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2780467 · Received October 10, 2012

Report

Report Number
2124215-2012-12957
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 19, 2012
Report Date
September 19, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

A BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DISCUSSED THE CLINICAL OBSERVATIONS WITH THE CALLER AND RECOMMENDED A CHEST X-RAY. THE CALLER STATED THE PATIENT IS DOING FINE NOW BUT A REVISION IS LIKELY TO OCCUR IN THE FUTURE. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING PACING IMPEDANCE MEASUREMENTS LESS THAN 200 OHMS. ADDITIONALLY, INTERMITTENT NOISE WAS OBSERVED DURING ISOMETRICS WHICH RESULTED IN SOME OVERSENSING BUT NO INAPPROPRIATE THERAPY WAS NOTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT A PROCEDURE OCCURRED TO SURGICALLY ABANDON AND REPLACE THIS LEAD. NO ADVERSE PATIENT EFFECTS OTHER THAN THE ADDITIONAL PROCEDURE WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0158

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| L| R E110| 4470| T125| 0158