CONTAK RENEWAL TR
Report
- Report Number
- 2124215-2012-12324
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- September 6, 2012
- Report Date
- April 21, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- DTB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE AND LEAD REMAIN IMPLANTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED AND AN AMENDED REPORT SUBMITTED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED REVEALED THAT ALL THE SEAL PLUGS WERE MISSING. THE MARKS IN THE HEADER INDICATE THE LV LEAD WAS FULLY INSERTED INTO THE HEADER.THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A FOLLOW UP VISIT, THIS LEFT VENTRICULAR (LV) LEAD AND CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXHIBITED LOSS OF CAPTURE AND HIGH IMPEDANCES ON THE LEFT VENTRICULAR CHANNEL WHEN PROGRAMMED USING THE LV TIP ELECTRODE. A CHEST X-RAY WAS PERFORMED AND REVEALED THAT THE LEAD HAD NOT MOVED. IN ADDITION, SATISFACTORY PARAMETERS WERE OBTAINED WHEN EITHER THE TIP OR RING WERE NOT INCLUDED IN THE PROGRAMMING VECTOR. A BOSTON SCIENTIFIC TECHNICAL SERVICE CONSULTANT REVIEWED THIS EVENT AND DISCUSSED THAT THESE OUT OF RANGE MEASUREMENTS COULD BE AN INDICATION OF A LEAD MECHANICAL ISSUE INVOLVING THE LEAD TIP CONDUCTOR OR THE CONNECTOR. IN ADDITION, THE DEVICE SETSCREW OF THE LEAD TIP MAY NOT BE FULLY TIGHTENED. IT WAS THEN SUSPECTED THAT THE LV LEAD HAD MOVED IN THE HEADER OF THE DEVICE, CAUSING OUT OF RANGE MEASUREMENTS IN CERTAIN PROGRAMMED VECTORS. THE DECISION WAS MADE TO LEAVE THE LEAD AND DEVICE IMPLANTED PROGRAMMED WITH AN OPTIMAL VECTOR. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
ADDITIONAL INFORMATION INDICATES THAT THIS PRODUCT WAS LATER EXPLANTED AND RETURNED FOR AN UNSPECIFIED REASON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL TR | IMPLANTABLE CHF GENERATOR | DTB | GUIDANT CRM CLONMEL IRELAND | H140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |