FDA Adverse Event Injury Summary report: N

LIVIAN

MDR report key: 2780095 · Received October 10, 2012

Report

Report Number
2124215-2012-12606
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 3, 2012
Report Date
August 28, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

THE PATIENT WAS SUBSEQUENTLY SEEN FOR TESTING. A MAXIMUM ENERGY SHOCK WAS DELIVERED AND SHOCK IMPEDANCE MEASUREMENTS WERE NOTED TO BE 40 OHMS. THE PHYSICIAN WILL CONTINUE TO MONITOR THE DEVICE/LEAD SYSTEM. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS ROUTINELY EXPLANTED AND REPLACED TWO YEARS LATER.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED A LOW OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIVIAN IMPLANTABLE CHF GENERATOR LWP GUIDANT CRM CLONMEL IRELAND H227

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L| R 0184| 4469| 4543| H227