FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 2780022 · Received October 10, 2012

Report

Report Number
2124215-2012-12737
Event Type
Injury
Date Received
October 10, 2012
Date of Event
September 10, 2012
Report Date
October 29, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. ON VISUAL INSPECTION, THERE WAS DRIED BLOOD OR BODY FLUID NOTED IN THE LEAD LUMEN. ALSO, ELECTROCAUTERY MELTED INSULATION WAS OBSERVED AT 79, 182, 203, 425, AND 675 MM FROM THE TERMINAL PIN; AT 835 MM FROM THE TERMINAL PIN THE INSULATION WAS BUNCHED. THE INSULATION ISSUES ARE CONSISTENT WITH INDUCED DAMAGE FROM THE EXPLANT PROCEDURE. RESISTANCE TESTING WAS COMPLETED TO ASSESS THE LEAD'S ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES THAT WOULD CAUSE A DISLODGEMENT.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS AVAILABLE, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH CAPTURE THRESHOLDS WITH LOSS OF CAPTURE NOTED. IT WAS DETERMINED THE LEAD WAS DISLODGED AND SUBSEQUENTLY SURGICAL INTERVENTION WAS PERFORMED. THIS LEAD WAS EXPLANTED AND A NEW LV LEAD WAS SUCCESSFULLY PLACED IN A DIFFERENT BRANCH. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUITY IMPLANTABLE HF LEAD OJX CPI - DEL CARIBE 4592

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| L| R 4542| 4136| 0181| N119| 4592