FDA Adverse Event Malfunction Summary report: N

ACUITY

MDR report key: 2780017 · Received October 10, 2012

Report

Report Number
2124215-2012-12059
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
July 30, 2012
Report Date
September 6, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE FOLLOW-UP, THIS LEFT VENTRICULAR (LV) LEAD WAS NOTED TO HAVE A PRODUCT PERFORMANCE ISSUE. WHEN A PACING VECTOR WAS USED THAT INCLUDED THE LV RING, OUT OF RANGE PARAMETERS WERE SEEN. PACING IMPEDANCE WAS HIGH AND OUT OF RANGE AND THERE WAS NO CAPTURE AT MAXIMUM OUTPUT. WHEN A DIFFERENT VECTOR WAS CHOSEN THAT UTILIZED THE LV TIP TO RV RING, THIS WAS FOUND TO HAVE NORMAL IMPEDANCE AS WELL AS CAPTURE IN THE LEFT VENTRICLE. NORMAL BI-VENTRICULAR PACING WAS ACHIEVED IN THIS VECTOR. THERE WAS NO REPORT OF NOISE ON THE EGM, AND AN X-RAY SHOWED NO SIGNS OF LEAD MOVEMENT. A BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) CONSULTANT WAS CONTACTED WITH REGARD TO THIS ISSUE. TS SAID THAT THE OBSERVATIONS SEEN WITH THIS LEAD COULD BE INDICATIVE THAT THE LEAD IS NOT OPTIMALLY INSERTED IN THE DEVICE HEADER (DEVICE MODEL/SERIAL P108/(B)(4)). IN SUCH SITUATION, THE TIP ELECTRODE COULD STILL MAKE CONTACT WITH THE CONNECTOR BLOCK. THIS WOULD ALLOW TO USE IT FOR PACE/SENSE CONFIGURATION (AS DESCRIBED IN THE HOSPITAL FORM: LV TIP TO RV RING/COIL). THE PATIENT WAS NOTED TO HAVE RARE DIAPHRAGMATIC TWITCH AT OUTPUTS ABOVE 3V @ 0.8MS, HOWEVER THIS SETTING PROVIDED THE BEST PARAMETERS, SO IT WAS LEFT AND THE PATIENT WAS ADVISED TO CONTACT THE CLINIC IF ANY ISSUES WERE OBSERVED. THERE IS NO PLAN TO REPLACE THIS LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUITY IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4555

Patients

Seq Age Sex Outcome Treatment
1