FDA Adverse Event Malfunction Summary report: N

GALILEO

MDR report key: 2775784 · Received October 8, 2012

Report

Report Number
1034569-2012-00184
Event Type
Malfunction
Date Received
October 8, 2012
Date of Event
September 8, 2012
Report Date
October 8, 2012
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
BK040013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

BLUD DIRECT WAS UTILIZED TO REVIEW THE IMAGE RESULTS. ALL TEST WELLS WERE CONSISTENT WITH THE REACTION STRENGTHS REPORTED BY THE INSTRUMENT. THE CUSTOMER INDICATED THAT THEIR SOP DOES NOT ALLOW FOR REPEAT TESTING ON THE INSTRUMENT. THE CUSTOMER ALSO DECLINED SERVICE DUE TO ALL OTHER SAMPLES ON THE SAME PLATE RESULTING AS EXPECTED. QC RESULTED AS EXPECTED. THERE WERE NO ERRORS REPORTED AT THE TIME OF TESTING. THE CUSTOMER SUBMITTED THREE TUBES FROM PATIENT SAMPLE FOR EVALUATION BY (B)(4) PRODUCT INVESTIGATIONS LABORATORY. MANUAL TUBE TESTING WAS PERFORMED USING RETENTIONS ANTI-B SERIES 3, LOT 203382 AND ANTI-B GAMMA CLONE, LOT 205022-2 WITH CUSTOMER'S SUBMITTED SAMPLES, (B)(4) (TUBES A, B, AND C). CUSTOMER'S SAMPLE EXHIBITED NEGATIVE REACTIVITY. A FWD_ABORH ASSAY WAS PERFORMED WITH CUSTOMERS SUBMITTED SAMPLES ON THE GALILEO USING ANTI-B SERIES 3, LOT 203382. QC PERFORMED AS EXPECTED. TUBE A: 1ST RUN RESULTED INV (INVALID) DUE TO INSTRUMENT ERROR. THE REPEAT RUN REPORTED FORWARD AS AN A POS, TUBE B FORWARD AS AN A POS, AND TUBE B WAS CLOTTED AND COULD NOT REPORT VALID RESULTS. ALL SAMPLES PERFORMED AS EXPECTED. THE REACTIVITY OF RETENTION ANTI-B SERIES 3, LOT 203382 DEMONSTRATED EXPECTED REACTIVITY.

Description of Event or Problem · 1

A CUSTOMER REPORTED OBTAINING AN ABO MISTYPE ON THE GALILEO WITH THE REFLEX ABO A, B ASSAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GALILEO AUTOMATED BLOOD BANK ANALYZER KSZ IMMUCOR, INC.

Patients

Seq Age Sex Outcome Treatment
1