GALILEO
Report
- Report Number
- 1034569-2012-00184
- Event Type
- Malfunction
- Date Received
- October 8, 2012
- Date of Event
- September 8, 2012
- Report Date
- October 8, 2012
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- BK040013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
BLUD DIRECT WAS UTILIZED TO REVIEW THE IMAGE RESULTS. ALL TEST WELLS WERE CONSISTENT WITH THE REACTION STRENGTHS REPORTED BY THE INSTRUMENT. THE CUSTOMER INDICATED THAT THEIR SOP DOES NOT ALLOW FOR REPEAT TESTING ON THE INSTRUMENT. THE CUSTOMER ALSO DECLINED SERVICE DUE TO ALL OTHER SAMPLES ON THE SAME PLATE RESULTING AS EXPECTED. QC RESULTED AS EXPECTED. THERE WERE NO ERRORS REPORTED AT THE TIME OF TESTING. THE CUSTOMER SUBMITTED THREE TUBES FROM PATIENT SAMPLE FOR EVALUATION BY (B)(4) PRODUCT INVESTIGATIONS LABORATORY. MANUAL TUBE TESTING WAS PERFORMED USING RETENTIONS ANTI-B SERIES 3, LOT 203382 AND ANTI-B GAMMA CLONE, LOT 205022-2 WITH CUSTOMER'S SUBMITTED SAMPLES, (B)(4) (TUBES A, B, AND C). CUSTOMER'S SAMPLE EXHIBITED NEGATIVE REACTIVITY. A FWD_ABORH ASSAY WAS PERFORMED WITH CUSTOMERS SUBMITTED SAMPLES ON THE GALILEO USING ANTI-B SERIES 3, LOT 203382. QC PERFORMED AS EXPECTED. TUBE A: 1ST RUN RESULTED INV (INVALID) DUE TO INSTRUMENT ERROR. THE REPEAT RUN REPORTED FORWARD AS AN A POS, TUBE B FORWARD AS AN A POS, AND TUBE B WAS CLOTTED AND COULD NOT REPORT VALID RESULTS. ALL SAMPLES PERFORMED AS EXPECTED. THE REACTIVITY OF RETENTION ANTI-B SERIES 3, LOT 203382 DEMONSTRATED EXPECTED REACTIVITY.
A CUSTOMER REPORTED OBTAINING AN ABO MISTYPE ON THE GALILEO WITH THE REFLEX ABO A, B ASSAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GALILEO | AUTOMATED BLOOD BANK ANALYZER | KSZ | IMMUCOR, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |