FDA Adverse Event Other Summary report: N

TELESCOPE, 10MM, 0 DEGREE, HD, AUTOCLAVABLE

MDR report key: 2773529 · Received September 26, 2012

Report

Report Number
9610773-2012-00083
Event Type
Other
Date Received
September 26, 2012
Date of Event
November 16, 2011
Report Date
November 16, 2012
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FBP
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE INVOLVED IN THE EVENT HAS NOT BEEN RETURNED TO MFR FOR INVESTIGATION, BUT HAS BEEN INSPECTED BY THE LOCAL SALES AND SERVICE ORGANIZATION WITHOUT OBSERVATION. AS THE PRODUCTS OF OTHER MANUFACTURERS TOO, THE VIDEOTELESCOPE HEATS UP AT THE DISTAL END DURING LIGHT EMISSION. THE HEATING-UP EFFECT OF THE WA53000A IS NEVERTHELESS IN CONFORMANCE TO THE REQUIREMENTS OF (B)(4). IT CAN BE CONSIDERED BASIC KNOWLEDGE FOR MEDICAL PROFESSIONALS TO SWITCH-OFF THE VIDEOTELESCOPE WHEN NOT IN USE, ESPECIALLY WHEN IT IS PUT ASIDE OR ON THE PT. FOR THAT PURPOSE AND TO MEET CUSTOMER EXPECTATIONS REGARDING USABILITY, THE PRODUCT IS EQUIPPED WITH A STAND-BY MODE. APPROPRIATE WARNING MESSAGE ARE INCLUDED IN THE IFU. THE EVENT IS AN OBVIOUS CONSEQUENCE OF A USE ERROR WITH RATHER MINOR (TEMPORARY) CONSEQUENCES. THE CASE WILL BE CLOSED WITHOUT FURTHER ACTIONS.

Description of Event or Problem · 1

THE PHYSICIAN PERFORMED A LAPAROSCOPIC OOPHOROCYSTECTOMY. DURING THE PROCEDURE, THE PT SUFFERED A SKIN BURN ON HER THIGH, AS THE DOCTOR HAD PUT THE DEVICE IN A POCKET OF THE DRAPE WITHOUT SWITCHING THE LIGHT OFF. THE DOCTOR CARRIED OUT TREATMENT OF THE BURNS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELESCOPE, 10MM, 0 DEGREE, HD, AUTOCLAVABLE OLYMPUS VIDEOTELESCOPE FBP OLYMPUS WINTER & IBE GMBH WA53000A

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization