FDA Adverse Event
Malfunction
Summary report: N
DUAL COOLER/HEATER
MDR report key: 2772262
·
Received October 2, 2012
Report
- Report Number
- 1828100-2012-01259
- Event Type
- Malfunction
- Date Received
- October 2, 2012
- Date of Event
- September 7, 2012
- Report Date
- September 7, 2012
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP
- Product Code
- DWC
- PMA / PMN Number
- PREAMED
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE USER FACILITY'S BIOMEDICAL ENGINEER WAS ABLE TO DETERMINE THE PROBABLE CAUSE WAS DEFECTIVE THERMOSTATS. REPLACEMENT PARTS WERE ORDERED BY THE CUSTOMER. THE DEFECTIVE CAN NOT BE RETURNED. IF ADD'L INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE OF THE DEVICE AT THE USER FACILITY'S SERVICE CENTER, THE HEATER COOLER WOULD NOT HOLD TEMPERATURE. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUAL COOLER/HEATER | COOLING & HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP | 11160 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |