FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 2772262 · Received October 2, 2012

Report

Report Number
1828100-2012-01259
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 7, 2012
Report Date
September 7, 2012
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP
Product Code
DWC
PMA / PMN Number
PREAMED
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE USER FACILITY'S BIOMEDICAL ENGINEER WAS ABLE TO DETERMINE THE PROBABLE CAUSE WAS DEFECTIVE THERMOSTATS. REPLACEMENT PARTS WERE ORDERED BY THE CUSTOMER. THE DEFECTIVE CAN NOT BE RETURNED. IF ADD'L INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE OF THE DEVICE AT THE USER FACILITY'S SERVICE CENTER, THE HEATER COOLER WOULD NOT HOLD TEMPERATURE. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUAL COOLER/HEATER COOLING & HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP 11160

Patients

Seq Age Sex Outcome Treatment
1