BARD PLATINUM PLATINUM CLASS URETEROSCOPIC 3
Report
- Report Number
- 1018233-2012-01437
- Event Type
- Malfunction
- Date Received
- October 2, 2012
- Report Date
- September 5, 2012
- Manufacturer
- OMNITECH SYSTEMS, INC.
- Product Code
- FFL
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE WAS RECEIVED AND THE BREAK WAS CLEAN AND ALL PIECES OF THE SHEATH APPEAR TO BE PRESENT. VISUAL EXAMINATION NOTED THAT THE ACTUAL BASKET WAS INTACT WITH NO SIGNS OF DAMAGE. NO OBVIOUS MANUFACTURING DEFECTS OR OTHER CONDITIONS WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE BROKEN SHEATH. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE REPORTED LOT NUMBER DID NOT SHOW ANY PROBLEMS OR CONDITIONS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. THE INSTRUCTIONS FOR USE WHICH ACCOMPANIES ALL DEVICES CURRENTLY ADDRESSES POTENTIAL RISKS ASSOCIATED WITH SURGICALLY IMPLANTED MATERIALS. THE LABELING AND INSTRUCTIONS FOR USE STATES: ¿CAUTIONS: OBJECTS THAT ARE TOO LARGE TO BE RECOVERED THROUGH THE SHEATH OR THROUGH THE SCOPE CHANNEL WILL REQUIRE THE SCOPE AND BASKET TO BE REMOVED SIMULTANEOUSLY FROM THE URINARY TRACT. IF RESISTANCE IS ENCOUNTERED DURING ADVANCEMENT OR WITHDRAWAL OF THE DEVICE, STOP AND DETERMINE THE SOURCE OF RESISTANCE, AS CONTINUED RESISTANCE MAY DAMAGE THE DEVICE AND COULD RESULT IN PATIENT INJURY. TAKE ACTION TO ALLEVIATE THE RESISTANCE. WHERE NECESSARY, USE OF A LITHOTRITE MAY BE REQUIRED TO REDUCE THE STONE BURDEN WITHIN THE BASKET, PROVIDED THAT NO DIRECT CONTACT IS MADE WITH THE STONE BASKET. PRECAUTIONS: DO NOT ALLOW THE DEVICE TO COME IN CONTACT WITH ANY ELECTRICAL EQUIPMENT. DO NOT ROTATE AN OPEN BASKET IN THE URETER. POTENTIAL COMPLICATIONS THAT MAY RESULT FROM THE USE OF A BASKET IN AN ENDOSCOPIC UROLOGICAL PROCEDURE INCLUDE BUT ARE NOT LIMITED TO: PERFORATION, EDEMA, EVULSION, ENTRAPMENT, BASKET INVERSION, INABILITY TO DISENGAGE FROM, HEMORRHAGE IRRETRIEVABLE OBJECT. (B)(4).
IT WAS REPORTED THAT THE BASKET TIP BROKE OFF IN THE PATIENT. THE TIP WAS REMOVED DURING THE PROCEDURE USING GRASPING FORCEPS THROUGH THE INITIALLY PLACED SCOPE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BARD PLATINUM PLATINUM CLASS URETEROSCOPIC 3 | FFL | OMNITECH SYSTEMS, INC. | NA | BMVKM029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |